ACCUCARETM
GAMMA-GT-SLR
Kinetic test (SCE recommendations)
Order Information
Catalog No: Product Description: Pack Size:
GGTSLR 10 GAMMA-GT 1 x 10 ml

METHOD
Kinetic test

PRINCIPLE
Gamma-glutamyl is transferred from gamma-glutamyl-p-nitroanilide to glycylglycine by gamma-GT (Gamma-Glutamyl-Transferase). The p-nitroanline formed absorbs at 405 nm. The amount of p-nitroani is directly proportional to gamma-GT activity.

SAMPLE:
Serum.

REFERENCE VALUES

REFERENCE VALUES
  WOMEN MEN
37°C 9 - 37 U/l 11 - 43 U/l

The reference values are only indicative in nature. Every laboratory should establish its own normal ranges.

QUALITY CONTROL :
Accutestrol N - H

REAGENTS
Reagent I : Buffer reagent
Reagent II : Substrate reagent

REFERENCES
Szasz G., Clin. Chem., (1969), 15: 24,

REAGENT PREPARATION
Mix 4 ml of Buffer Reagent with 1 ml of Enzyme Reagent.

STABILITY: The working reagent is stable for 30 days at 2-8°C.

AUTOMATED PARAMETERS
Wave length 405 nm
Cuvette 1 cm light path
Reaction Temperature 37°C
Measurement Against distilled water
Reaction Type Kinetic test
Reaction Direction Increasing
Sample/Reagent Ratio 1: 10
Delay/Lag/time 60 Secs
Interval time 30 Secs
No of Readings 04
Blank absorbance limit <0.85
Factor 1111
Low Normal at 37°C 9 U/L
High Normal at 37°C 43 U/L
Linearity at 37°C 200 U/L

MANUAL PROCEDURE

PIPETTE INTO TEST TUBE

SAMPLE 100 µl
REAGENT 1000 µl

Mix well and incubate for 1 min. at 37°C. read initial absorbance and start timer simultaneously; read again after 1,2 and 3 min (p A/min).

CALCULATION AND LINEARITY
p A/min. x 1111 = U/l Gamma GT

The method is linear upto 200 U/l.