ORDER
INFORMATION
| Catalog No: |
Product Description: |
Pack Size: |
| CAOC 100 |
CALCIUM |
2 X 50 ML |
| CAOC 200 |
CALCIUM |
2 X 100 ML |
METHOD
Colorimetric end point test
PRINCIPLE
Calcium ions form a violet complex with O-cresolphthalein complexone
in an alkaline medium.
SAMPLE
Serum, heparinised plasma, 24 urine diluted 1:3.
REFERENCE VALUES
| SERUM |
9 - 11 mg/dl |
| 24 h URINE |
50 - 400 mg/24h |
The reference values are to be considered as indicative only. Every
laboratory should establish its own normal ranges.
QUALITY CONTROL
Accutestrol N - H
REAGENTS
Reagent I : AMP Buffer reagent
Reagent II : OCPC Color reagent
Reagent III : Calcium standard : 10 mg/dl
REFERENCES Faulker W.R, and Meites. S., Caly J.P. et. al gitelman,
H. anal, Richterich. R, 59, 836. |
REAGENT PREPARATION
Mix equal volumes of Reagent I and Reagent II to give enough reagent
for the number of samples being run.
STABILITY
6 hrs. at room temperature
AUTOMATED PARAMETERS
| Wave length |
570 nm |
| Cuvette |
1 cm |
| Measurement |
Against reagent blank |
| Reaction Temperature |
R.T. |
| Reaction Type |
End Point |
| Sample/Reagent Ratio |
1:40 |
| Incubation |
5 minutes |
| Reagent Blank Abs limit |
<0.5 |
| Low Normal |
9.0 mg/dl |
| High Normal |
11 mg/dl |
| Linearity |
15 mg/dl |
MANUAL PROCEDURE
PIPETTE INTO TEST TUBE
| |
BLANK |
STD |
SAMPLE |
| SAMPLE |
- |
- |
25 µl |
| STANDARD |
- |
25 µl |
- |
| REAGENT |
1000 µl |
1000 µl |
1000 µl |
Mix well, incubate at R.T. 5 mins. Measure final absorbance of the
sample (Ac) and standard (As) against the reagent blank.
CALCULATION AND LINEARITY
| Ac/As x C = mg/dl SERUM |
| Ac/As x C x 3 = mg/dl URINE |
C = Concentration Standard
The method is linear to a concentration of 15 mg/dl. |