ORDER
INFORMATION
| Catalog No: |
Product Description: |
Pack Size: |
| CARZM 50 |
CALCIUM |
50 X 1 ML |
| CARZ 100 |
CALCIUM |
2 X 50 ML |
| CARZ 200 |
CALCIUM |
2 X 100 ML |
METHOD
Colorimetric end point test
PRINCIPLE
At a neutral pH, the Ca2+ forms with arsenazo III a complex, the color
intensity of which is directly proportional to the concentration of
calcium in the sample.
SAMPLE
Serum, heparinised plasma, 24 urine diluted 1:3 with distilled water
(adjust pH 3-4 with 0.1 N HCL).
REFERENCE VALUES
The reference values are to be considered as indicative only. Every
laboratory should establish its own normal ranges..
QUALITY CONTROL
Accutestrol N - H
REAGENTS
Reagent I : Arsenazo III reagent
Reagent II : Calcium standard : 10 mg/dl
NOTE:
CARZM 50 are specially pre-treated monovials with 1 ML pre-dispensed
reagent. Just add 25 µl sample / std. & aspirate. Use the
same programme.
REFERENCES
Bishop,M.L.Duben-Von Laufen, J.L,.Burtis, Carl Aa and Ashwood.
|
REAGENT PREPARATION
All reagents are ready to use
STABILITY : Upto expiry date when stored at R.T.
AUTOMATED PARAMETERS
| Wave length |
650 nm |
| Cuvette |
1 cm light path |
| Measurement |
Against reagent blank |
| Reaction Temperature |
R.T. |
| Reaction Type |
End Point |
| Sample/Reagent Ratio |
1:40 |
| Incubation |
5 minutes |
| Blank absorbance limit |
0.80 |
| Low Normal |
8.8 mg/dl |
| High Normal |
10.2 mg/dl |
| Linearity |
16 mg/dl |
MANUAL PROCEDURE
PIPETTE INTO TEST TUBE
| |
BLANK |
STD |
SAMPLE |
| SAMPLE |
- |
- |
25 µl |
| STANDARD |
- |
25 µl |
- |
| REAGENT |
1000 µl |
1000 µl |
1000 µl |
Mix well, incubate at R.T. 5 mins. Measure final absorbance of
the sample (Ac) and standard (As) against the reagent blank.
CALCULATION AND LINEARITY
| Ac/As x C = mg/dl SERUM |
| Ac/As x C x 3 = mg/dl URINE |
C = Concentration Standard