ORDER
INFORMATION
| Catalog No: |
Product Description: |
Pack Size: |
| LDH 10 |
LDH-L |
10 X 1.1 ML |
| LDH 30 |
LDH-L |
10 X 3 ML |
METHOD
Kinetic test
PRINCIPLE
Lactate is oxidized to Pyruvate in the presence of NAD by the action
of lactate dehydrogenase. The rate of formation of NADH is directly
proportional to LDH concentration.
SAMPLE
Serum free of hemolysis
REFERENCE VALUES
| Males |
50-166 U/L 30°C
80-285 U/L 37°C |
| Females |
60-132 U/L 30°C
103-227 U/L 37°C |
QUALITY CONTROL :
Accutestrol N - H
REAGENTS
Reagent I : Enzyme reagent
REFERENCES
Teitz, R.W. Henry, J.B., Henry, R.J.Young, D.S. 123.
REAGENT PREPARATION
Dissolve contents of each vial of enzyme reagent with the volume of
deionised water as specified on the vial.
|
STABILITY: 14 days when stored at 2-8°C (protected from
light)
AUTOMATED PARAMETERS
| Wave length |
340 nm |
| Cuvette |
1 cm light path |
| Reaction Temperature |
37°C |
| Measurement |
Against distilled water |
| Reaction Type |
Kinetic test |
| Reaction Direction |
Increasing |
| Sample/Reagent Ratio |
1: 40 |
| Delay/Lag/time |
60 Secs |
| Interval time |
30 Secs |
| No of Readings |
04 |
| Blank absorbance limit |
<0.6 |
| Factor |
6592 |
| Low Normal at 37°C |
80 U/L |
| High Normal at 37°C |
285 U/L |
| Linearity at 37°C |
800 U/L |
MANUAL PROCEDURE
PIPETTE INTO TEST TUBE
| Working Reagent |
1000 µl |
| Sample |
25 µl |
Mix well and after 1 min incubation, measure the change in absorbance
per minute (p A/min.) during 2 minutes.
CALCULATION AND LINEARITY
| 340
nm : Activity (U/L) = p A/min
X 6592 |
Linearity at 37°C is 800 U/L |