ACCUCARETM
PHOSPHORUS UV
End-point method
|
ORDER
INFORMATION
| Catalog No: |
Product Description: |
Pack Size: |
| PHOSM 50 |
PHOSPHORUS
(MONO) |
50 X 1 ML |
| PHOS 100 |
PHOSPHORUS |
2 X 50 ML |
| PHOS 200 |
PHOSPHORUS |
2 X 100 ML |
METHOD
UV endpoint test
PRINCIPLE
Inorganic phosphate reacts in acid environment with molybdic
acid to form an unreduced phosphomolybdic acid complex, which absorbs
light at 340 nm. The absorbance is directly proportional to the phosphorus
concentration in the sample.
SAMPLE:
Serum. Plasma, 24h Urine diluted 1:20 with distilled water.
REFERENCE VALUES
| SERUM |
|
| Men |
2.1 - 5.6 mg/dl |
| Women |
1.5 - 6.8 mg/dl |
| Children |
4.0 - 7.0 mg/dl |
| 24 h Urine |
0.3 - 1.0 mg/dl |
The reference values are only indicative in nature. Every laboratory
should establish its own normal ranges.
QUALITY CONTROL :
Accutestrol N - H
REAGENTS
Reagent I : Molybdate reagent
Reagent II : Phosphorus standard 5.0 mg/dl
REFERENCES
Teitz, N.W., Fundamentals of Clinical Chemistry, 2nd edition,
W.B.Saunders, Philadelphia, 1976
|
REAGENT PREPARATION
All reagent are ready to use.
STABILITY : Upto expiry date when stored at 2-8°C.
AUTOMATED PROCEDURE
| Wave length |
340 nm |
| Reaction Type |
End Point |
| Cuvette |
1 cm light path |
| Reaction Temperature |
R.T. |
| Reaction Direction |
Increasing |
| Measurement |
Against reagent blank |
| Incubation |
10 minutes |
| Maximum absorbance blank |
<0.9 |
| Low Normal |
2.1 mg/dl |
| High Normal |
5.6 mg/dl |
| Linearity |
12.0 mg/dl |
MANUAL PROCEDURE
PIPETTE INTO TEST TUBE
| |
BLANK |
STD |
SAMPLE |
| Sample |
|
|
20 µl |
| STANDARD |
|
20 µl |
|
| REAGENT |
1000 µl |
1000 µl |
1000 µl |
Mix well and incubate at RT for 10 minutes. Measure final absorbance
of the sample (Ac) and standard (As) against the reagent blank.
CALCULATION AND LINEARITY
Ac
/ As x Conc. Std = mg / dl Serum
Ac / As x Conc. Std x 20 = mg / dl urine |
Where Vdl is the urine volume in dl
The method is linear to a concentration of 12.0 mg/dl. |
|
 |
Lab
Care Diagnostics (India) Pvt. Ltd.
16-A, 'A' Wing , 1st Floor, Sita Estate, Mahul
Road, Aziz Baug, Chembur, Mumbai 400 074. India
Phone: 00 91 22 2554 2109/ 00 91 22 2554 1558 Fax: 00 91 22 2554 3541
E-mail : accucare@labcarediagnostics.com |
|