ORDER
INFORMATION
| Catalog No: |
Product Description: |
Pack Size: |
| GGT 10 |
GAMMA - GT |
10 x 1.1 ml |
| GGT 30 |
GAMMA - GT |
10 x 3 ml |
METHOD
Kinetic test
PRINCIPLE
Gamma-glutamyl is transferred from gamma-glutamyl-p-nitroanilide
to glycylglycine by gamma-GT (Gamma-Glutamyl-Transferase). The p-nitroanine
formed absorbs at 405 nm. The amount of p-nitroanine is directly
proportional to gamma-GT activity.
SAMPLE
Serum.
REFERENCE VALUES
| |
WOMEN |
MEN |
| 37°C |
9 - 37 U/l |
11 - 43 U/l |
The reference values are only indicative in nature. Every laboratory
should establish its own normal ranges.
QUALITY CONTROL :
Accutestrol N - H
REAGENTS
Reagent I : Buffer reagent
Reagent II : Substrate reagent
REFERENCES
Szasz G., Clin. Chem., (1969), 15: 24,
REAGENT PREPARATION
Dissolve the contents of one vial of substrate reagent with volume
of buffer reagent as specified on the vial.
|
STABILITY: 21 days at 2-8°C.
AUTOMATED PARAMETERS
| Wave length |
405 nm |
| Cuvette |
1 cm light path |
| Reaction Temperature |
37°C |
| Measurement |
Against distilled water |
| Reaction Type |
Kinetic test |
| Reaction Direction |
Increasing |
| Sample/Reagent Ratio |
1 : 10 |
| Delay/Lag/time |
60 Secs |
| Interval time |
30 Secs |
| No: of Readings |
04 |
| Blank Absorbance limit |
<0.85 |
| Factor |
1111 |
| Low Normal at 37°C |
9 U/L |
| High Normal at 37°C |
43 U/L |
| Linearity at 37°C |
200 U/L |
MANUAL PROCEDURE
PIPETTE INTO TEST TUBE
| STANDARD |
100 µl |
| REAGENT |
1000 µl |
Mix well and incubate for 1 min. at 37°C. read initial absorbance
and start timer simultaneously; read again after 1,2 and 3 min (p
A/min).
CALCULATION AND LINEARITY
| p
A/min. x 1111 = U/l Gamma GT |
The method is linear upto 200 U/L |