ACCUCARETM
TOTAL PROTEIN
Colorimetric test - Biuret
|
ORDER
INFORMATION
| Catalog No: |
Product Description: |
Pack Size: |
| TP 100 |
TOTAL PROTEIN |
2 X 50 ML |
| TP 200 |
TOTAL PROTEIN |
2 X 100 ML |
METHOD
Colorimetric endpoint test
PRINCIPLE
Cupric ions, in an alkaline medium, interact with protein peptide
bonds resulting in the formation of a coloured complex.
REAGENTS
Reagent I : Biuret Reagent
Protein Standrad: 6 g/ dl (store at 2-8°C)
SAMPLE:
Serum. Heparinised plasma or EDTA plasma.
REFERENCE VALUES
6.6 - 8.3 g/dl
QUALITY CONTROL
Accutestrol N - H
REFERENCES
Gournall A et al Journal of Biol. Chem 177 (1949), 751. Tietz, N.W.,
Fundamentals of Clinical Chemistry. W.B. Saunders Co., Philadelphia,
PA (1970), 302. |
REAGENT PREPARATION
The reagent are ready to use.
STABILITY: upto expiry date when stored at R.T.
AUTOMATED PARAMETERS
| Wave length |
540 NM |
| Reaction Temperature |
20 - 25°C |
| Reaction Type |
End Point |
| Cuvette |
1 cm light path |
| Measurement |
Against reagent blank |
| Sample/Reagent Ratio |
1: 50 |
| Incubation |
10 minutes |
| Low Normal |
6.6 g/dl |
| High Normal |
8.3 g/dl |
| Linearity |
10 g/dl |
MANUAL PROCEDURE
PIPETTE INTO TEST TUBE
| |
BLANK |
STD |
SAMPLE |
| SAMPLE |
- |
- |
20 µl |
| STANDARD |
- |
20 µl |
- |
| REAGENT |
1000 µl |
1000 µl |
1000 µl |
Mix well, incubate for 10 minutes at 20-25°C.
Measure absorbance of the sample (Ac) and
standard (As) against reagent blank.
CALCULATION AND LINEARITY
| Ac
/ As X conc. Std. = g/dl PROTEIN |
The method is linear to a concentration of 10 g/dl.
If the concentration of the sample exceeds 10 g/dl,
dilute 1+1 with 0.9% saline solution. Multily the result by 2. |
|
 |
Lab
Care Diagnostics (India) Pvt. Ltd.
16-A, 'A' Wing , 1st Floor, Sita Estate, Mahul
Road, Aziz Baug, Chembur, Mumbai 400 074. India
Phone: 00 91 22 2554 2109/ 00 91 22 2554 1558 Fax: 00 91 22 2554 3541
E-mail : accucare@labcarediagnostics.com |
|